Boston Scientific Corporation CRM70514 Benutzerhandbuch

Seite von 86
38
Q-TECH™ PROGRAMMER:  OPERATiOn
4. 
Select the On/Off Therapy switch to set the desired therapy mode (Figure 30).
Figure 30:  Setting the On/Off Therapy switch
5. 
Select and drag the Conditional Shock Zone (yellow) and Shock Zone (red) slider bars 
to set the desired zone configuration.
• 
The Shock Zone is programmable between 170 and 250 bpm in steps of 10 bpm.
• 
The Conditional Shock Zone is programmable between 170 and 240 bpm in steps 
of 10 bpm. Enhanced detection criteria are automatically enabled when the Condi-
tional Shock Zone is programmed.
• 
When programming both the Shock Zone and Conditional Shock Zone, maintain 
at least a 10 bpm difference between the two zones. If the Conditional Shock Zone 
slider (yellow) is dragged over the Shock Zone slider (red), the two sliders will merge 
to create a single Shock Zone.
6. 
If post-shock pacing is desired, slide the Post Shock Pacing switch to the On position. 
(Post-shock bradycardia pacing occurs at a non-programmable rate of 50 bpm for up 
to 30 seconds. Pacing is inhibited if the intrinsic rate is greater than 50 bpm.)
7. 
Select the Program button to apply the changes and program the SQ-RX device. 
A screen will appear to confirm that the SQ-RX device settings were programmed 
(Figure 31). If the SQ-RX device does not accept the SQ-RX device programming, a 
Program Device alert screen will appear. Select the Try Again button to return to the 
Device Settings screen.