ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
SpO
2
 measurement may be incorrect in the 
following cases.
•  When the patient’s carboxyhemoglobin or 
methemoglobin increases abnormally.
•  When dye is injected in the blood.
• During CPR.
•  When measuring at a site with venous pulse.
•  When there is body movement.
•  When the pulse wave is small (insufficient 
peripheral circulation).
WARNING
Check the circulation condition by observing the 
skin color at the measurement site and pulse 
waveform. Change the measurement site every 8 
hours for disposable probes and every 4 hours for 
reusable probes (every 8 hours for TL-631T series 
probe). The skin temperature may increase at the 
attached site by 2 or 3°C (4 or 5°F) and cause a 
burn or pressure necrosis. When using the probe 
on the following patients, take extreme care and 
change the measurement site more frequently 
according to symptoms and degree.
•  Patient with a fever
•  Patient with insufficient peripheral circulation
•  Neonate or low birth weight infant with delicate 
skin
WARNING
When monitoring SpO
2
 of a patient who is 
receiving photodynamic therapy, the light from the 
finger probe sensor may cause a burn. 
Photodynamic therapy uses a photosensitizing 
agent that has a side effect of photosensitivity.