ZM-540PA, ZM-541PA Operator’s Manual
When the probe is attached on an appropriate site
with sufficient thickness and the error message
confirming the probe attachment repeatedly
appears, the probe may be deteriorated. Replace it
with a new one.
with sufficient thickness and the error message
confirming the probe attachment repeatedly
appears, the probe may be deteriorated. Replace it
with a new one.
CAUTION
When a message indicates a faulty probe, stop
monitoring and replace the probe with a new one.
monitoring and replace the probe with a new one.
CAUTION
While a patient is on medication which causes
vasodilation, the pulse waveform may change and
in rare cases the SpO
vasodilation, the pulse waveform may change and
in rare cases the SpO
2
value might not be
displayed.
CAUTION
When monitoring SpO
2
only, detection of
arrhythmia and asystole is not available and
arrhythmia alarms such as ASYSTOLE, VF or VT
are not available. If the patient requires ECG
monitoring, monitor the ECG.
arrhythmia alarms such as ASYSTOLE, VF or VT
are not available. If the patient requires ECG
monitoring, monitor the ECG.
CAUTION
When monitoring SpO
2
only (without ECG
monitoring), turn on both the upper and lower limit
alarms for PR and SpO
alarms for PR and SpO
2
on the receiving monitor. If
the patient’s pulse is not detected during asystole
or other condition, a “CANNOT DETECT PULSE”
or “CHECK PROBE” alarm occurs instead of an
SpO
or other condition, a “CANNOT DETECT PULSE”
or “CHECK PROBE” alarm occurs instead of an
SpO
2
limit alarm. Furthermore, if the patient has no
pulse, noise from probe movement could be
misjudged as a pulse and cause an incorrect PR or
SpO
misjudged as a pulse and cause an incorrect PR or
SpO
2
value to be displayed.
CAUTION