Nihon Kohden Corporation ZM-541PAA Benutzerhandbuch
18
ZM-540PA, ZM-541PA Operator’s Manual
Only use Nihon Kohden specified electrodes and
electrode leads. When other type of electrodes or
electrode leads are used, the “CHECK
ELECTRODES” message may be displayed and
ECG monitoring may stop.
electrode leads. When other type of electrodes or
electrode leads are used, the “CHECK
ELECTRODES” message may be displayed and
ECG monitoring may stop.
CAUTION
When the “CHECK ELECTRODES” message is
displayed on the receiving monitor, ECG is not
monitored properly and the ECG alarm does not
function. Check the electrode, electrode leads, and
if necessary, replace with new ones.
displayed on the receiving monitor, ECG is not
monitored properly and the ECG alarm does not
function. Check the electrode, electrode leads, and
if necessary, replace with new ones.
CAUTION
SpO
2
Monitoring
WARNING
SpO
2
measurement may be incorrect in the
following cases.
• When the patient’s carboxyhemoglobin or
• When the patient’s carboxyhemoglobin or
methemoglobin increases abnormally.
• When dye is injected in the blood.
• During CPR.
• When measuring at a site with venous pulse.
• When there is body movement.
• When the pulse wave is small (insufficient
• During CPR.
• When measuring at a site with venous pulse.
• When there is body movement.
• When the pulse wave is small (insufficient
peripheral circulation).
WARNING
When monitoring SpO
2
of a patient who is
receiving photodynamic therapy, the light from the
finger probe sensor may cause a burn.
Photodynamic therapy uses a photosensitizing
agent that has a side effect of photosensitivity.
finger probe sensor may cause a burn.
Photodynamic therapy uses a photosensitizing
agent that has a side effect of photosensitivity.
7KH6S2
2
SUREHVPDQXIDFWXUHGE\1LKRQ.RKGHQKDYHWZRZDYHOHQJWKVZLWKSHDNVLQWKH
UDQJHRIDQGQP7KHPD[LPXPOLJKWLQWHQVLW\LVOHVVWKDQP: