ZM-540PA, ZM-541PA Operator’s Manual
WARNING
Check the circulation condition by observing the
skin color at the measurement site and pulse
waveform. Change the measurement site every 8
hours for disposable probes and every 4 hours for
reusable probes (every 8 hours for TL-631T series
probe). The skin temperature may increase at the
attached site by 2 or 3°C (4 or 5°F) and cause a
burn or pressure necrosis. When using the probe
on the following patients, take extreme care and
change the measurement site more frequently
according to symptoms and degree.
• Patient with a fever
• Patient with insufficient peripheral circulation
• Neonate or low birth weight infant with delicate
skin color at the measurement site and pulse
waveform. Change the measurement site every 8
hours for disposable probes and every 4 hours for
reusable probes (every 8 hours for TL-631T series
probe). The skin temperature may increase at the
attached site by 2 or 3°C (4 or 5°F) and cause a
burn or pressure necrosis. When using the probe
on the following patients, take extreme care and
change the measurement site more frequently
according to symptoms and degree.
• Patient with a fever
• Patient with insufficient peripheral circulation
• Neonate or low birth weight infant with delicate
skin
The disposable probe is not sterilized. Use the
disposable probe only for a single patient. Never
reuse the disposable probe for another patient
because it causes cross infection.
disposable probe only for a single patient. Never
reuse the disposable probe for another patient
because it causes cross infection.
CAUTION
If the attachment site is dirty with blood or bodily
fluids, clean the attachment site before attaching
the probe. If there is nail polish on the attachment
site, remove the polish. Otherwise, the amount of
transmitted light decreases, and measured value
may be incorrect or measurement cannot be
performed.
fluids, clean the attachment site before attaching
the probe. If there is nail polish on the attachment
site, remove the polish. Otherwise, the amount of
transmitted light decreases, and measured value
may be incorrect or measurement cannot be
performed.
CAUTION
If the skin gets irritated or redness appears on the
skin from the probe, change the attachment site or
stop using the probe. Take extreme care for the
patients with delicate skin.
skin from the probe, change the attachment site or
stop using the probe. Take extreme care for the
patients with delicate skin.
CAUTION