ZM-540PA, ZM-541PA  Operator’s Manual
Only use Nihon Kohden specified electrodes and 
electrode leads. When other type of electrodes or 
electrode leads are used, the “CHECK 
ELECTRODES” message may be displayed and 
ECG monitoring may stop.
CAUTION
When the “CHECK ELECTRODES” message is 
displayed on the receiving monitor, ECG is not 
monitored properly and the ECG alarm does not 
function. Check the electrode, electrode leads, and 
if necessary, replace with new ones.
CAUTION
SpO
2
 Monitoring
WARNING
SpO
2
 measurement may be incorrect in the 
following cases.
•  When the patient’s carboxyhemoglobin or 
methemoglobin increases abnormally.
•  When dye is injected in the blood.
• During CPR.
•  When measuring at a site with venous pulse.
•  When there is body movement.
•  When the pulse wave is small (insufficient 
peripheral circulation).
WARNING
When monitoring SpO
2
 of a patient who is 
receiving photodynamic therapy, the light from the 
finger probe sensor may cause a burn. 
Photodynamic therapy uses a photosensitizing 
agent that has a side effect of photosensitivity.