ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
Check the circulation condition by observing the 
skin color at the measurement site and pulse 
waveform. Change the measurement site every 8 
hours for disposable probes and every 4 hours for 
reusable probes (every 8 hours for TL-631T series 
probe). The skin temperature may increase at the 
attached site by 2 or 3°C (4 or 5°F) and cause a 
burn or pressure necrosis. When using the probe 
on the following patients, take extreme care and 
change the measurement site more frequently 
according to symptoms and degree.
•  Patient with a fever
•  Patient with insufficient peripheral circulation
•  Neonate or low birth weight infant with delicate 
skin
The disposable probe is not sterilized. Use the 
disposable probe only for a single patient. Never 
reuse the disposable probe for another patient 
because it causes cross infection.
CAUTION
If the attachment site is dirty with blood or bodily 
fluids, clean the attachment site before attaching 
the probe. If there is nail polish on the attachment 
site, remove the polish. Otherwise, the amount of 
transmitted light decreases, and measured value 
may be incorrect or measurement cannot be 
performed.
CAUTION
If the skin gets irritated or redness appears on the 
skin from the probe, change the attachment site or 
stop using the probe. Take extreme care for the 
patients with delicate skin.
CAUTION