Operator's Manual  ZM-920PA/930PA
Front Panel
CAUTION
Only use your finger to press the CALL key.  Do not press the key
with a sharp object, otherwise the key may be damaged.
Power switch
Turns transmitter power on or off.
CALL key
When this key is pressed,
a "peep" sounds at the 
transmitter, and "CALL" 
message appears at the 
monitor.  Depending on the 
settings on the monitor, an 
ECG waveform is recorded 
when this key is pressed.
Channel lable
Replace batteries LED
Lights when the batteries 
need replacement.
Check electrodes LED
Lights when the electrode is 
detached from the patient.
Battery case
Contains two 1.5 V dry cell batteries (AA).
Refer to the warning below.
RA LA LL Va Vb
USAch
9002
608.025 MHz
WARNING
Close the battery case cover during operation.  If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed.  Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.
ZM-920PA