Nihon Kohden Corporation ZM-930PA Benutzerhandbuch

Seite von 70
Operator's Manual  ZM-920PA/930PA
27
2.
Carefully remove the backing paper from
the electrode.  Avoid touching the adhesive
surface.
3.
Place the electrode on the previously
cleaned skin.  Pay attention to the electrode
lead color and symbol.
4.
Fasten the electrode lead wire with surgical
tape with an extra length of wire between
the tape and the electrode.  This lessens the
movement of electrode leads by body
movement and helps stable monitoring.
Detection and Display of Measurement Condition
Electrode Detachment
The “
” LED lights on the transmitter or the “CHECK ELECTRODE” message
is displayed on the screen of the monitor in the following cases.
• Electrode is detached from skin.
• Electrode lead is disconnected from the electrode.
• Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is
displayed on the receiving monitor, ECG is not monitored properly
and the ECG alarm dose not function. Check the electrode and
electrode leads, and if necessary, replace with new ones.