Operator's Manual  ZM-920PA/930PA
29
Respiration Measurement Procedure
1.
Select the electrode lead and disposable electrodes.
2.
Connect the electrode lead to the ECG/RESP socket.
3.
Connect disposable electrodes to the electrode lead and attach electrodes to
the patient.
After steps 1 to 3 are finished, respiration monitoring automatically starts.
Electrode Position for Respiration Monitoring
Place the RA and LL electrodes so that the lungs are between the electrodes.
NOTE
The optimum electrode positions for ECG measurement of a patient
are not always optimum for respiration measurement of the patient.
Select positions suitable for both ECG and respiration
measurements, or positions which have priority for one
measurement.
Electrode Position Examples
NOTE
The following examples are when monitoring with 3 electrodes.  ECG
cannot be monitored correctly when electrodes are attached as the
following examples when monitoring with 6 electrodes.
Position 1
In this position, respiration measurement is available; however, there is a
difference in amplitude between different patients.
RA
 
LL
 
Right infraclavicular 
fossa 
Fifth intercostal space on the 
left midclavicular line, V4