Operator's Manual  ZM-920PA/930PA
Problem Cause Countermeasure 
3 electrode leads 
are properly 
attached to the 
patient but   
LEDs light. 
The transmitter is fixed to 
6 lead monitoring. 
Turn off and on the 
transmitter power. 
Heart rate of the 
patient who is 
using an electric 
blanket cannot be 
monitored on the 
receiving monitor. 
Pacing pulse detection is 
set to ON on the monitor. 
Turn off the pacing pulse 
detection. 
The gel on the electrode is 
dried out. 
The gel on the electrode is 
coming off. 
Replace the electrode with 
a new one. 
Electric blanket is used. 
Cover the blanket with a 
shield cover. 
ECG baseline is 
thick. 
(Hum is 
overlapping) 
Hum filter is set to OFF on 
the monitor 
Set the filter to ON. 
The gel on the electrode is 
dried out. 
Respiration 
waveform 
measurement is 
unstable. 
The gel on the electrode is 
coming off. 
Replace the electrode with 
a new one. 
The probe size is not 
appropriate for the patient.
Use the appropriate probe 
for the patient. 
Probe is attached to the 
same limb that is used for 
NIBP measurement. 
Attach the probe to the 
opposite limb. Avoid a site 
where blood circulation 
condition changes greatly. 
Probe attachment 
condition is poor. Probe is 
about to detach from the 
skin. External light gets in.
Firmly attach the probe 
according to the procedure 
in the probe operator’s 
manual. 
SpO
2
 data is 
unstable and not 
reliable. 
Measurement site is dirty. 
Patient is wearing nail 
polish. 
Remove dirt and nail 
polish.