Troubleshooting
If the problem still remains after checking the following, contact your Nihon
Kohden distributor.
Problem Cause  Countermeasure 
Batteries are not 
installed correctly. 
The battery 
polarity is wrong. 
Install the batteries correctly. 
The power cannot be 
turned on. 
Batteries are 
completely 
discharged. 
Replace the batteries with new 
ones. 
Nothing is displayed 
on the LCD after 
turning the power on. 
(ZM-930PA only) 
SpO
2
 display is 
turned off. 
Turn off the power, and turn on 
the power again. 
Nothing is displayed 
on the monitor after 
turning the 
transmitter power on. 
The channel of the 
transmitter and 
monitor does not 
match. 
Set the correct channel on the 
monitor. 
Electrode lead is 
not connected to 
the transmitter. 
Connect the electrode lead to the 
transmitter. ECG electrode lead 
works as an antenna for 
transmitting data to the receiving 
monitor. If ECG is not measured, 
the signal may not be received. 
Another 
transmitter of the 
same channel is 
used near by. 
Turn the transmitter power off. If 
the monitor still receives a signal, 
there is a high probability that 
another transmitter of the same 
channel is used nearby. 
Follow the instruction of your 
channel administrator and use 
another transmitter of a different 
channel. 
Signals are 
mixing. 
Follow the instruction of your 
channel administrator and use 
another transmitter of a different 
channel. 
Signal receiving 
condition is poor. 
The transmitter is 
damaged. 
Contact your Nihon Kohden 
distributor.