Operator's Manual ZM-920PA/930PA
Sound Display
Cause Countermeasure
Intermittent
“peep” sound
every 1 s
“peep” sound
every 1 s
(3)
(4) all lit
(5) blinking
(4) all lit
(5) blinking
Cannot receive
sufficient light from
the probe photo
emitter.
sufficient light from
the probe photo
emitter.
Attach probe to a
site with 6 to 14
mm thickness
where sufficient
light can be
received.
site with 6 to 14
mm thickness
where sufficient
light can be
received.
Intermittent
“pip” sound
every 0.25 s
“pip” sound
every 0.25 s
(3) 85%,
blinking
blinking
Broken probe
Replace the probe
with a new one.
with a new one.
Patient body
movement
movement
Remove the cause
by checking the
patient condition
and changing the
attachment site.
by checking the
patient condition
and changing the
attachment site.
--- (3)
blinking
Probe is not attached
securely.
securely.
Securely attach the
probe.
probe.