Display and Message List
Indication Cause
Countermeasure
Fully charged battery
Batteries are low.
Batteries are low. NIBP cannot be measured.
No
indication
Dead batteries
Replace batteries.
Battery Indication
ECG/Respiration
Indication Cause
Countermeasure
Electrode lead is disconnected from the
electrode.
electrode.
Firmly connect the electrode
lead to the electrode.
lead to the electrode.
Electrode lead is disconnected from the
transmitter.
transmitter.
Firmly connect the electrode
lead to the transmitter.
lead to the transmitter.
Electrode lead discontinuity.
Replace the electrode lead
with a new one.
with a new one.
Electrode is not firmly attached to the skin.
Polarization voltage is abnormally high.
Replace the electrode with a
new one.
new one.
When monitoring ECG with 6 electrodes, the electrode or lead detached position is indicated by
either lead or electrode position. This is set at LEADS OFF DISPLAY on the PARAMETER
SETUP screen. Refer to the “Changing Parameter Setup Settings” section.
either lead or electrode position. This is set at LEADS OFF DISPLAY on the PARAMETER
SETUP screen. Refer to the “Changing Parameter Setup Settings” section.
LEADS OFF DISPLAY set to CHAR
ECG ELECTRODE set to AHA
ECG ELECTRODE set to AHA
LEADS OFF DISPLAY set to CHAR
ECG ELECTRODE set to IEC
ECG ELECTRODE set to IEC
LEADS OFF DISPLAY set to IMAGE