Display and Message List
Indication Cause 
Countermeasure 
 
 
Fully charged battery 
 
 
 
Batteries are low.   
 
 
Batteries are low. NIBP cannot be measured. 
No 
indication 
Dead batteries 
Replace batteries. 
 
Battery Indication
ECG/Respiration
Indication Cause 
Countermeasure 
Electrode lead is disconnected from the 
electrode. 
Firmly connect the electrode 
lead to the electrode. 
Electrode lead is disconnected from the 
transmitter. 
Firmly connect the electrode 
lead to the transmitter. 
Electrode lead discontinuity. 
Replace the electrode lead 
with a new one. 
Electrode is not firmly attached to the skin. 
 
Polarization voltage is abnormally high. 
Replace the electrode with a 
new one. 
 
When monitoring ECG with 6 electrodes, the electrode or lead detached position is indicated by
either lead or electrode position.  This is set at LEADS OFF DISPLAY on the PARAMETER
SETUP screen.  Refer to the “Changing Parameter Setup Settings” section.
LEADS OFF DISPLAY set to CHAR
ECG ELECTRODE set to AHA
LEADS OFF DISPLAY set to CHAR
ECG ELECTRODE set to IEC
LEADS OFF DISPLAY set to IMAGE