NIBP
Message Cause 
Countermeasure 
The cuff and extension hose are not 
properly connected. 
The cuff hose (or extension hose) is 
not properly connected to the NIBP 
socket. 
Connect them properly. 
NIBP AIR LEAK 
The cuff or extension hose is 
damaged. 
Replace with a new one. 
The patient’s pulse wave is small. 
Measure by palpation or auscultation. 
NIBP CANNOT 
DETECT PULSE 
The cuff is not wrapped on the 
patient properly. 
Wrap the cuff on the patient properly. 
NIBP CHANGE 
BATTERIES NO NIBP 
NIBP cannot be measured due to 
low battery. 
Replace batteries with new ones. 
NIBP CUFF 
OCCLUSION 
Transmitter malfunction. 
Immediately remove the cuff from the 
patient and contact your Nihon Kohden 
representative. 
NIBP HIGH CUFF 
PRESS 
Enormous pressure was applied by 
the pressure of the cuff. 
Remove the cause. 
NIBP INFLATION 
PRESS LOW 
Insufficient cuff inflation pressure. 
Wait for the remeasurement to be 
performed with increased cuff inflation 
pressure. 
NIBP MEAS 
TIME-OUT 
The measuring time exceeded the 
specified time due to arrhythmia, 
body movement, vibration or, cuff 
or air hose being squeezed. 
Remove the cause if the cause is body 
movement, vibration or squeezing of 
cuff or hose. 
NIBP MODULE 
FAILURE 
Module malfunction. 
Contact your Nihon Kohden 
representative. 
NIBP REMEASURING 
NIBP is being remeasured due to 
arrhythmia, body movement, 
vibration or, cuff or air hose being 
squeezed. 
If the message still appears after 
remeasurement, remove the cause if 
the cause is body movement, vibration 
or squeezing of cuff or hose. 
Check that the hose is not bent or 
squeezed. 
NIBP SAFETY 
CIRCUIT RUNNING 
(When this message is 
displayed, measurement 
cannot be performed for 
40 seconds.) 
Measurement stopped by the safety 
circuit. 
Wait 40 seconds, then perform 
remeasurement. If the message still 
appears, contact your Nihon Kohden 
representative. 
NIBP SYS OUT OF 
RANGE 
The maximum blood pressure 
cannot be measured even when the 
cuff inflation pressure exceeded 280 
mmHg when using adult cuff. 
Measure by palpation or auscultation. 
The patient’s pulse wave is too 
small. 
Measure by palpation or auscultation. 
The cuff is wrapped too loosely. 
Wrap the cuff properly. 
NIBP WEAK PULSE 
The cuff size is not appropriate. 
Use the appropriate cuff. 
NIBP ZEROING 
NIBP zero balance is being 
adjusted. 
Do not touch the cuff during zeroing. 
Wait for the message to disappear.