Operator's Manual  ZM-940PA
Message Cause  Countermeasure 
During NIBP 
measurement 
SpO
2
 monitoring is paused for 
NIBP measurement. 
Wait for NIBP measurement to 
finish. 
Considerable body movement. 
 Detecting 
body 
movement 
The probe is not attached to 
the patient properly. 
When the message is displayed 
frequently, check the patient 
condition and, if necessary, change 
the attachment site. 
The probe is not attached to 
the patient properly. 
Attach the probe to the patient 
properly. 
The probe is not attached at 
the appropriate site. 
Attach the probe to a site 6 to 14 
mm thick. 
SpO
2
 CHECK 
PROBE 
Probe is expired. 
Replace the probe with a new one. 
Searching for the correct pulse 
wave. 
Wait until the pulse wave is 
detected. 
The SpO
2
 value cannot be 
obtained because the 
waveform is unstable. 
SpO
2
 DETECTING 
PULSE 
The probe is not attached to 
the patient properly. 
Attach the probe to the patient 
properly. 
SpO
2
 LIGHT 
INTERFERENCE 
SpO
2
 measurement site is 
under fluorescent light, 
surgical light, sunlight, etc. 
Cover the measurement site with a 
blanket or cloth. 
Probe is expired. 
Replace the probe with a new one. 
SpO
2
 PROBE 
FAILURE 
Probe is damaged or 
short-circuited. 
Replace the probe with a new one. 
Poor peripheral circulation. 
Check the patient condition and 
change the attachment site. 
SpO
2
 WEAK PULSE  The probe is attached too 
tightly and is obstructing the 
blood circulation. 
Check the probe attachment 
condition and if necessary, 
reattach the probe. 
 
SpO
2