Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
REMOTE CONTROL 
115 
SYSTem:BUZZer:FTIMe 
Set
Query
 
 
Description 
Sets the FAIL Sound duration in seconds.  
Syntax 
Query Syntax 
SYSTem:BUZZer:FTIMe <NR2> 
SYSTem:BUZZer:FTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.2~999.9 
Example 
SYST:BUZZ:FTIM 1 
Sets the buzzer to 1 second for a FAIL judgment. 
SYSTem:ERRor 
Query
 
 
Description 
Returns any errors in the output buffer. See page 
143 for a list of all the possible errors. 
Query Syntax 
SYSTem:ERRor ? 
Return parameter  <string> 
Returns an error string. 
Example 
SYST:ERR ? 
>Value Error! 
Returns “Value Error!” as the error message. 
SYSTem:GPIB:VERSion 
Query
 
 
Description 
Queries the GPIB version.  
Query Syntax 
SYSTem:GPIB:VERSion? 
Return parameter  <string> 
Returns: 
The GPIB version as a string 
“GPIB,V1.00”  
or  
“No GPIB connected” if there is not a 
GPIB device configured/connected. 
Query Example 
SYST:GPIB:VERS? 
>GPIB,V1.00 
Returns the GPIB version.