Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
116
 
Function Commands 
 
 
FUNCtion:TEST ...................................................................................... 116
 
MEASure<x> .......................................................................................... 117
 
MAIN:FUNCtion .................................................................................... 118
 
 
FUNCtion:TEST 
Set
Query
 
 
Description 
Turns the currently selected test (output) on or off.  
 
When HOLD is displayed on the screen during 
AUTO tests, use the FUNCtion:TEST command to 
move on to the next step.  
 
Setting the FUNCtion:TEST command to OFF at 
the end of a test will also temporarily turn the 
PASS/FAIL buzzer sound off.  
Syntax 
Query Syntax 
FUNCtion:TEST {ON|OFF} 
FUNCtion:TEST? 
Parameter 
ON 
Turns the test on. 
OFF 
Turns the test off. 
Return parameter  TEST ON 
Test is on. 
TEST OFF 
Test is off. 
Example 
FUNC:TEST ON 
Turns the output on.