117 
MEASure<x> 
Query
 
 
Description 
Returns the test parameters & results of the tester 
in either MANU or AUTO mode.  
MANU mode: Returns the test parameters & 
results of a MANU test. 
AUTO mode: Returns the test parameters & results 
of the selected step (1-16) of the AUTO test.  
Return parameters: function, judgment/status, test 
voltage, test current/resistance, test time (time of 
completed test) or ramp time (elapsed time of test 
that has not been completed.    
Query Syntax 
MEASure<x>? 
Parameter 
(MANU mode) 
 
No parameter needed for MANU 
mode. 
Parameter  
(AUTO mode) 
<x> 
<NR1>1~16. Step number.  
Return parameter  <string> 
Returns the test status of the test 
in the following format: 
function, judgment or status, test 
voltage, test current or resistance, 
test time or ramp time 
 
 
Function 
ACW, DCW, IR, GB 
Judgment 
/Status 
PASS, FAIL 
VIEW 
Test voltage 
voltage+unit 
Test current 
/Test resistance 
current+unit 
resistance+unit 
Test time 
/Ramp time 
T=time+S 
R=time+S  
Example  
(in MANU mode) 
MEAS? 
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S 