The following error messages or messages may appear on the GPT 
screen when configuring the MANU test settings.   
 
Error Messages 
Description 
TIME ERR 
For ACW tests. TIME ERR is displayed when HI 
SET ≥ 30.00mA~40.00mA and if the RAMP  
time and the TEST TIME setting is > 240 
seconds. 
OVER 50W 
For DCW tests. OVER 50W is displayed if the 
HI SET setting multiplied by the Voltage setting 
is greater than 50W. 
I ERR 
For ACW, DCW tests. Shown when the current 
is set too high. 
SHORT 
Voltage is too low. Indicates that the DUT 
could be shorted. 
V ERR 
For ACW, DCW tests. Voltage is too high. 
V = 0 
For GB tests. Voltage is equal to 0. Check to 
see that the SENSE H or SOURCE H is not 
open. 
R ERR 
For IR tests. The voltage is too high or 
resistance=0Ω. Check to see whether the DUT 
or test lead is shorting.  
 
For GB tests. The resistance is too high or too 
low. Check the test lead connections to correct 
the error. 
I<SET 
For GB tests. Current too low. Indicates that 
the SOURCE L or SOURCE H test lead is open 
or poorly connected. Test the test lead 
connection with the DUT to confirm. 
I>SET 
For GB tests. Current is too high. 
R=0 
For GB tests. Resistance = 0. This error 
indicates that there is an error with the 
measured resistance (0Ω). Perform the zeroing 
function again.