APPENDIX
149
GPT-9000 Specifications
The specifications apply when the GPT-9000 is powered on for at
least 30 minutes at 15˚C~35˚C.
least 30 minutes at 15˚C~35˚C.
Specifications
Environment
Range
Temperature
Humidity
Warranty
15˚C ~ 35˚C
≤70% (No
condensation)
condensation)
Operation
0˚C ~ 40˚C
≤70% (No
condensation)
condensation)
Storage
-10˚C ~ 70˚C
≤85% (No
condensation)
condensation)
Installation Location
Indoors at an amplitude of up to 2000m.
AC Withstanding Voltage
Output Voltage Range
0.100kV~ 5.000kV
Output Voltage Resolution
2V
Output Voltage Accuracy
(1% of setting +5V) with no load
Maximum Rated Load(Table1) 200 VA (5kV/40mA)
Maximum Rated Current
Maximum Rated Current
40mA
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 40mA(0.5kV<V≤5kV)
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 40mA(0.5kV<V≤5kV)
Output Voltage Waveform
Sine wave
Frequency
50 Hz / 60 Hz
Voltage Regulation
1% +5V
[Maximum rated load → no load]
Voltmeter Accuracy
(1% of reading+ 5V)
Current Measurement Range 0.001mA~040.0mA
Current Best Resolution
Current Best Resolution
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)
Current Measurement
Accuracy
Accuracy
± (1.5% of rdg + 30 counts) when HI SET<1.00mA
± (1.5% of rdg + 3 counts) when HI SET≥1.00mA
± (1.5% of rdg + 3 counts) when HI SET≥1.00mA
Window Comparator Method Yes
ARC DETECT
ARC DETECT
Yes
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
Continued…