Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
150
 
TIMER (Test Time) 
OFF*, 0.5S~999.9S 
GND 
ON/OFF 
* The timer can only be turned off under special MANU mode (MANU=***-000) 
 
DC Withstanding Voltage 
Output Voltage Range 
0.100kV~ 6.000kV 
Output Voltage Resolution 
2V 
Output Voltage Accuracy 
 (1% of setting +5V) with no load 
Maximum Rated Load(Table1)  50W (5kV/10mA) 
Maximum Rated Current 
10mA 
0.001mA ~ 2mA (0.1kV≤V≤0.5kV) 
0.001mA ~ 10mA (0.5kV<V≤6kV) 
Voltmeter Accuracy 
 (1% of reading+ 5V) 
Voltage Regulation 
 1% +5V 
[Maximum rated load → no load] 
Current Measurement Range  0.001mA~010.0mA 
Current Best Resolution 
1uA 
0.001mA(0.001mA~0.999mA) 
0.01mA(01.00mA~09.99mA) 
0.1mA(010.0mA) 
Current Measurement 
Accuracy 
± (1.5% of reading +30uA) 
Window Comparator Method  Yes 
ARC DETECT 
Yes 
Rise-time Control Function 
Yes 
RAMP (Ramp Time) 
0.1~999.9S 
TIMER (Test Time) 
OFF*, 0.5S~999.9S 
GND 
ON/OFF 
* The timer can only be turned off under special MANU mode (MANU=***-000) 
 
Insulation Resistance Test 
Output Voltage 
50V~1000V 
Output Voltage Resolution 
50V 
Output Voltage Accuracy 
(1% of setting+5V) with no load 
Resistance Measurement 
Range 
1M
~ 9500M
 
Test Voltage 
Measurement Range  Accuracy 
50V≤V<500V 
1~50MΩ 
±(5% of reading +1MΩ) 
51~2000MΩ
 
±(10% of reading +1MΩ) 
500V≤V≤1000V 
1~500MΩ
 
±(5% of reading +1MΩ) 
501~9500MΩ
 
±(10% of reading +1MΩ) 
Voltage Regulation 
±1% +5V [Maximum rated load → no load] 
Window Comparator Method  Yes 
Rise-time Control Function 
Yes 
Continued…