8.8  Protective conductor current (I
PE
)
Current that flows through the protective conductor from the devices (test
objects) of protection class I, when their bodies are insulated from earth.
A direct measurement is applicable if the test object can be insulated from
earth.
In all other cases the differential current method should be applied, see Point
6.3 (Basic settings of the tester, Page 10).
Also see measuring principle circuits on Page 18.
Caution! The test object is supplied with mains voltage
during the test.
8.9  Protective conductor resistance (R
PE
)
Resistance between any conductive touchable parts, connected to the
protective conductor for protective purposes, and the earthed contact of the
mains plug, the device plug or the protective conductor, which is constantly
connected to the mains power supply.
During the measurement of the protective conductor resistance, the
connecting cable should be moved, section by section, along its whole
length.  This measuring method only applies for devices of protection class I.
Also see measuring principle circuit M1 on Page 17.
8.10 Visual inspection
Testing in accordance with DIN VDE 0701 or 0702 also involves visual
inspection of the device.
According to the standard the devices should be inspected for external
defects (without opening the device) and, as far as possible, for suitability to
the place of installation. In particular, attention should be paid to the
following:
– Damage to the housing
– Defects of the cord connector guard and strain relief fittings
– External damage of the connecting cables
– Unauthorised interference and changes
– Signs of overloading and improper use
– Orderly condition of the protective covers
– Safety impairing soiling and corrosion
– Unblocked cooling vents
– Existence of necessary air filters
– Leak tightness, pressure relief valve
– Legibility of safety-related inscriptions
– Fuse links in accordance with manufacturer’s specifications
Externally visible defects, which could lead to a mechanical or fire
hazard should be repaired immediately.