9.4a Measuring the protective conductor current
 (with devices of protection class I)
The limit value is 3.5 mA.
In the case of test objects with heating elements with a total capacity
greater than 3.5 kW, the protective conductor current may not be
greater than 1 mA/kW heat output. The protective conductor current can
either be measured directly with the substitute leakage current method
or with the differential current method.
Protective conductor current – differential current method PC I, Diagram M6
Measuring principle circuits
Protective conductor current – substitute leakage current method PC I, Diagram M4a
Tester
Test object
Protective conductor current - direct measurement PC I, Diagram M5
Tester
Test object
Caution! The test object is supplied with mains voltage
during the test.
Tester
Test object
18
Operating manual for the TG uni 1 appliance tester
Conducting tests in accordance with VDE 0701 and 0702 - definition of standards
Test socket
Test socket
Test socket
Insulated installation
In case of test objects in PC I with touchable conductive parts that
are not connected to the protective conductor, an additional touch
current measurement must be carried out in line with PC II
(see Diagram 1 on Page 20)
&
&
&
&
&