COMMNAND OVERVIEW 
71 
Query example 
:CURRent:VA? 
>1.0A 
Returns the “A Value” current setting. 
:CURRent:VB 
Set
Query
 
 
Description 
Sets or queries the CC mode “B Value” (static 
mode) current. This command is applicable to 
static modes. 
Note: A different current value can be set for each 
current range (High/Mid/Low).   
Syntax 
:CURRent:VB {<NRf>[ A ] | MINimum | MAXimum } 
Query Syntax 
:CURRent:VB? 
Parameter  
<NRf>[A] 
“B Value” current value 
 
MINIMUM or MIN 
Minimum current level 
 
MAXIMUM or MAX 
Maximum current level 
Return parameter  <NR2> 
Returns the “B Value” 
current value. 
Example 
:CURRent:VB MIN 
Sets the current value to the minimum. 
Query example 
:CURRent:VB? 
>1.0A 
Returns the “B Value” current setting. 
:CURRent:SRATe 
Set
Query
 
 
Description 
Sets or queries the current slew rate for CC static 
mode.   
Syntax 
:CURRent:SRATe {<NRf>| MINimum | MAXimum } 
Query Syntax 
:CURRent:SRATe? 
Parameter  
<NRf> 
Sets the slew rate in 
mA/us 
 
MINIMUM or MIN 
Set to the highest slew rate 
 
MAXIMUM or MAX 
Set to the lowest slew rate