8.11 Resetting to Factory Default Parameters
For various reasons, you might wish to return the PT878 to its original 
settings. The Factory Defaults option enables you to return the meter to its 
preprogrammed default settings. To enter the option, scroll down to the 
Factory Defaults entry in the Service Menu and press 
[ENTER]
. The screen 
Note:
All sites and logs will be lost if you use the Factory Defaults option. 
Be sure to transfer or print all sites and logs before resetting!
Figure 129: Factory Defaults Option
1.
 To restore Factory Default settings:
Press 
[F2]
 (Cancel) to return to Operate Mode without changing the 
current settings.
Press 
[F3]
 (OK) to confirm that you wish to restore the factory default 
settings.
2.
The program asks for confirmation: “Are you SURE?” Repeat the 
procedure shown in Step 1 above.
The PT878 replaces the default site file and the global meter settings with 
the factory default settings, and returns to Operate Mode.
Factory Defaults
Cancel
OK
*** WARNING ***
You will LOSE all changes by
pressing OK.
Choose CANCEL to exit or OK to
proceed.