8.12 Updating PT878 Software
By using a Windows-based PC with an infrared adapter, PT878 users can 
update the meter’s operating software without the need to change the 
hardware. This software can include the bootloader, FPGA software (timing 
software for transit-time), instrument software, the meter string file and the 
Help string file. The updating procedure involves three steps:
1.
Setting up the PT878 in Flash Update mode.
2.
Selecting the loading interface.
3.
Loading the new software from the PC into the PT878.
CAUTION!
To guard against mishap, you should print out, download 
or otherwise record all logs, settings and other data you 
wish to save. 
IMPORTANT:
When uploading new software, the PT878 uses a 
“bootloader” program. Be sure that the software you are 
installing is compatible with the installed bootloader 
version, shown in the “About” option discussed on 
page 110.
 If you need to replace both programs, install the 
new bootloader 
first, before any other software.
With the current bootloader software (version 3.0 and later), two interfaces 
are available for sending software to the PT878:
IrOBEX, a recently introduced Microsoft infrared standard which allows 
more straightforward use and faster file transfers, while tolerating IR 
interruptions better.
IrCOMM, an infrared standard supported by certain Microsoft operating 
systems.
GE recommends updating software via the IrOBEX standard; however, the 
IrCOMM standard is available for users who have problems with IrOBEX. 