Setting The “ZERO” Point
3. After pressing the 
   key,  the display will flash
    “2  PT  CAL” followed by “00  PLATE”. An audible
    “beep” will also be heard. Place the probe on the bare 
    metal part and press until the shell contacts the surface.
4. The display will show “CAL” on the left and a 
    thickness reading (near zero) on the right. Remove the 
    probe from the bare metal surface.
5. The display will then show “MORE/SEL” and will 
    “beep”.
    a. At this point you can continue to take additional 
        “ZERO” measurements to use an averaged ZERO 
       calibration. This is recommended when the surface 
       of the bare metal part is rough or uneven.
    b. Otherwise press the 
ON
 key to continue.
Setting the “Calibration” (2nd) Point
6. The display will flash “W/SHIM” and continue to 
    “beep”.  Place the plastic shim on the same bare metal 
    and measure the shim’s thickness.
7. The display will show “ADJC” along with a thickness 
    reading. Note the reading and determine if it matches 
    the thickness value marked on the shim.
    If the displayed reading does not match the shim’s 
    thickness, use the 
or 
keys to increase or
    decrease the displayed value until it matches the value 
    marked on the plastic shim.
8. When the displayed value was adjusted to the value 
    marked on the shim, press the 
ON 
key. The display 
    will  flash “COMPLETE” and then “-- -- --”.
The calibration is now complete and you are ready to 
take measurements.