Arizona MAX-5000XL Benutzerhandbuch

Seite von 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 26 of 92 
The possible displays are: 
  CHAMBER STABLE - The oven has been at a stable temperature for at 
least the time period set. 
  CHAMBER STABILIZING:  LOW TEMP - Oven is below temperature 
setpoint. 
  CHAMBER STABILIZING:  HIGH TEMP - Oven is above temperature 
setpoint. 
  CHAMBER STABILIZING:  xx:xx - Oven temperature is within the correct 
range and the time remaining until specified time is displayed. 
  Ash Rate 
– This is used for high test temperatures that will ignite a sample.  
YOU MUST USE THIS TO PREVENT HEATER DAMAGE.   
 
The Ash Rate value is a rate of moisture loss used to limit the rate of increase 
in oven temperature to slow ignition and prevent flaming.  The normal 
temperature setpoint is limited to keep the moisture loss at less than the set ash 
rate.  For example, if an ash rate of 7 %/Minute is used, the temperature 
increase is cont
rolled to maintain a ≤ 7 %/Minute moisture loss during the test. 
  Start Temperature Control - Prevents a test from starting if the chamber 
temperature is more than two degrees hotter or colder than the IDLE 
TEMPERATURE.  This feature can improve the standard deviation of results 
of some temperature-sensitive materials.  
 
Ending Criteria 
The criteria used by the instrument to 
automatically end a test.  This feature 
makes it unnecessary for the operator 
to monitor the analyzer while it is 
performing a test.  Selecting this 
option will display the ending criteria 
options. 
 
  Prediction - The MAX 5000XL 
uses an advanced version of 
the AZI moisture prediction 
algorithm to automatically end 
each test early to speed test 
results.  The prediction method 
relies on the fact that most 
samples exhibit a weight loss 
trend such that the rate of 
volatilization is proportional to 
the amount of volatile material 
remaining in the sample.  
TO PREVENT SAMPLE IGNITION AND POSSIBLE HEATER DAMAGE, 
USE ASH RATE, NOT TEMP RATE, AT ASHING TEMPERATURES.  TEMP 
RATE WILL NOT PREVENT SAMPLE IGNITION.