Fluke Lawn Mower 2635A Benutzerhandbuch

Seite von 268
Memory Card Operations
Setup File Procedures
3
3-9
Setup File Procedures
Perform the following procedures to LOAD, STORE, and ERASE memory card
instrument configuration (SEtUP) files.
Using Setup Store
Perform the procedure in Figure 3-5 to save the current instrument configuration. The
communication parameters: baud, parity, CTS, and echo, remain as set previously by the
front panel controls or computer interface. The instrument automatically assigns the next
sequential SEtxx file name. When SEt99 is reached, the instrument loops back to reuse
previously assigned file names that have been erased or skipped over. To assign your
own file name, use the up/down and left/right arrow keys when creating the file.
FILES
ENTER
SEtUP
StorE
ENTER
ENTER
Typical
Required
Selecting the SETUP mode.  Press the FILES 
key to access the FILES menu.  Press the 
up/down arrow keys until SEtUP is displayed, 
then press the ENTER key.  The menu changes 
to SEtUP.  If an error message appears, see 
Table 3-1.
Selecting the STORE mode.  Press the 
up/down arrow keys until StorE is displayed in 
the SEtUP menu, then press the ENTER key.  
The menu changes to StorE and a file name is 
displayed.
Storing a SEtxx File.  Record the displayed 
SEtxx file name, where xx represents an 
instrument-assigned number between 00 and 99 
or use the up/down, left/right arrow keys to select 
the xx number.  Press the ENTER key to store 
the file.
Init
StAt
dir
dAtA
SEtUP
ErASE
StorE
LOAd
SEt99
SEtxx
SEt00
yES
no
SUrE
Overwriting a SEtxx File.  If the xx number 
selected in the previous step is already assigned, 
the existing file will be overwritten with the new 
file.  If this is desired, select yES then press the 
ENTER key.  If no is selected, the procedure will 
terminate.  If an error message appears, see 
Table 3-1.
FILES
ENTER
op41f.eps
Figure 3-5. Using SETUP STORE to Save Configuration Files