LogTag Recorders LogTag Analyzer Temperature Recorder Benutzerhandbuch

Seite von 140
 
111 
 
C
H A P T E R  
1 0  
Appendix 
In This Chapter 
Troubleshooting .....................................................111
Resolving USB Driver Problems ...............................113
Sensor Responsiveness...........................................136
 
 
Troubleshooting 
Problem 
Solution 
My LogTag appears to be “dead”. No 
lights are flashing. 
This could represent either of two conditions: 
1  The LogTag is ready to start, but has not 
been started yet. 
2  The LogTag is simply in a state where the 
configuration has the LED flashing turned 
off. This is an option that may be selected 
when a LogTag has been shipped to the 
user for the first time (to preserve battery 
life) or it may be that the last configuration 
setting selected no LED display. 
3  The LogTag is in "Hibernation" mode. Refer 
to "
Prolonging battery life
" for further 
information. 
4  The battery is dead and the LogTag is 
expired. 
How to test:  
1  Try pressing the START MARK button on the 
LogTag. 
2  Start the LogTag Analyzer software and 
place the LogTag into the Interface Cradle. 
Download the LogTag. If the LogTag can be 
identified by the software, then the battery 
is good, and the LogTag is OK.