Section 3: Electrical installation
3-9000-759 Rev D
June 2014
Conduit seals
 45
Figure 3-2  External ground lug 
3.4
Conduit seals
Conduit seals are required for meter installations in hazardous environments. Adhere to safety 
instructions to protect personnel and equipment.  
A.
A. External ground lug
EXPLOSION HAZARD
To reduce the risk of an explosion or fire, conduit runs must have a sealing fitting connected 
within 457.2 mm (18 inches) of the enclosure. Substitution of components may impair 
intrinsic safety of the meter.
Failure to keep covers tight during operation may result in serious injury or death.
EXPLOSION HAZARD
Substitution of components may impair the intrinsic safety and cause ignition of flammable 
or combustible atmospheres. Disconnect power before servicing.
Failure to remove power and use Daniel approved components  may cause serious injury.