A&D AD-4405 User Manual

Page of 80
AD-4405 Weighing Indicator
 Page 
11 
 
 
5.
5.
5.
5.    Calibration
Calibration
Calibration
Calibration    
 
This indicator, converts an input voltage from a load cell to the "mass" value, and 
displays it. Calibration is the adjustment function so that the scale (indicator) can 
display the weight correctly. 
 
5.1.1. 
Items of Calibration Mode 
 
There are four items in the calibration function where setting should be done. 
 
How to calibrate: In the weighing mode, press the CAL key. After Cal in is displayed for 
2 seconds Cal 0 will appear. Then the required items should be selected and displayed 
with the 
MODE
 key, then executed by pressing the 
ENTER
 key.   
NOTE: Calibration could be started by simultaneously pressing the 
ZERO
 and 
TARE
 keys, 
instead of the CAL key.  However, in some cases, the procedures are altered to nullify 
this option.    Furthermore, this option is not available with sealed type version setting. 
Required Items 
Cal5et 
Store capacity, resolution, alignment of decimal point position and display 
format, weighing range and unit. These items should be set first in order 
for the indicator to function as a weighing instrument. Set values do not 
need to be changed again unless the indicator itself is replaced. For 
details, refer to “5.2.1. Setting a Weighing Instrument”. 
Cal 0 
Calibrates zero and span. This is required after installation, to get accurate 
data. For details, refer to “5.2.3. Zero Calibration” and “5.2.4. Span 
Calibration”. 
Optional Items (Sub-functions) 
lnr 0 
Performs digital linearization. Refer to “5.4. Digital Linearization 
Function”. 
g 5et 
Compensates for acceleration of gravity.
 Refer to “5.5. Gravity 
Compensation Function”. 
Gravity compensation function: Compensates for weighing error 
between the calibration location and another weighing location using 
gravity acceleration. 
 
In the calibration mode the keys have functions as follows: 
0
 - 
9
 Numerical 
keys. 
MODE
 
The key to display other items. 
CLEAR
 
The cancel key at inputting data, initial data and changing mode. 
+/- 
The key to display other parameters. 
ESC
 
The key to proceed to the next step without changing set values. 
ENTER
 
The key to store new calibration data and proceeds to the next step. 
CAL
 
The key to store all of the parameters into memory and display Caloff after 
the calibration mode. Press the 
ON/OFF
 key to turn off the display. 
  
5. 
Calibration