AD-4405 Weighing Indicator
 Page 
13 
 
 
Dual Range 
<First range> Select the resolution, decimal point position and format. 
¯ 
<First range> Select the weighing range 
¯ 
<Second range> Select the resolution 
¯ 
<Second Range> Specify the weighing capacity 
 
For the range function, refer to “5.3. Weighing Range Function”. 
 
 
 
Specifying 
Specifying 
Specifying 
Specifying the Range and Unit
the Range and Unit
the Range and Unit
the Range and Unit    
Step  1 
The range and unit of measure are displayed. 
Range display : 
5ingl : single range 
 
dUal : dual range 
To change the range function, use the 
CLEAR
 key 
Unit display: 
The active unit is displayed. When using units that can be 
used interchangeably, such as kg or lb, calibration should be 
done by using the displayed unit. 
 
To select a unit for weighing, use the 
MODE
 key and to select unit for 
calibration, use the +/- key. The unit for calibration (first unit) is 
displayed and the alternate unit (second unit) is blinking. 
ENTER
 
Stores the value displayed and then advances to the next step. 
ESC
 
Advances to the next step without changing the parameter. 
 
 
 
Specifying 
Specifying 
Specifying 
Specifying the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format    
Step  2  The resolution will be displayed as d 0.1, with decimal point. Triangle 1 and the first 
unit selected at the previous step will be displayed. 
Positioning of the decimal point is done with the 
CLEAR
 key, setting of display 
format (point or comma) with the +/- key, the resolution with the 
MODE
 key. The 
decimal point format, set at this stage, will only apply to the display. The decimal 
point format for serial data output is selected using the F-function settings. 
Press the 
ENTER
 key to store the displayed settings and then proceed to the next 
step. Using the 
ESC
 key, regardless of what is displayed, the indicator will 
proceed to the next step without changing the set data. 
 
 
 
 
 
5.2. Calibration Procedure