A&D AD-4405 User Manual

Page of 80
AD-4405 Weighing Indicator
 Page 
69 
 
 
19.
19.
19.
19.     4
4
4
4-
-
-
-20mA Analog Output (OP
20mA Analog Output (OP
20mA Analog Output (OP
20mA Analog Output (OP-
-
-
-07)
07)
07)
07)    
  The OP-07 analog output option is for sending the weight data to an analog input unit. 
  The output is a 4mA to 20mA current output proportional to the display reading. 
  The output data is updated in synchronization with the display update. 
 
Specifications
Specifications
Specifications
Specifications    
Output current 
4mA to 20mA    *
Non-linearity 
Less than +/- 0.1% fs 
Load resistance  0 to 510 
Temperature 
ZERO
Less than +/- 0.02% fs/°C
Resolution Approx. 
1/10000
 
coefficient 
SPAN
Less than +/- 0.02% fs/°C
Output terminal  Connector terminal      No.1 : +    No.2 : -    No.3 : FG(Earth) 
Accessory 
Connector terminal (1)    TM-MSTB03STF 
AC adapter (1) 
Note:  Please confirm that the AC adapter type is correct 
for your local voltage and receptacle type.
 
* When set to non-weight display (Calibration, 
F-settings etc.), output current is 4mA. 
  The output current is not adjustable. 
 
Settings
Settings
Settings
Settings    
Set f30 1 of the F-Function, and set f31, f32 and f33. 
 
Item 
Parameter
Remarks 
f30 
* 0 
No output 
Initial setting 
 Data output 
Analog output 
Must be set to 1 
 
Serial in/out 1 
 
Data 
output 
 
Serial in/out 2 
 
f31 
* 0 
Displayed value
Initial setting 
 Output data 
GROSS weight 
 
 
NET weight 
 
f32 
Weight value at 4mA 
-999999 to 999999 
(Initial setting is 0) 
Decimal point is set 
at “CALSEt” 
Analog 
output 
f33 
Weight value at 20mA 
-999999 to 999999 
(Initial setting is 10000) 
Decimal point is set 
at “CALSEt” 
Settings of f32 and f33 
When entering the f32 or f33 settings (press the 
ENTER
 key when “32” or “33” is blinking), the 
setting value is displayed.  Set the value using 
the 
0
 - 
9
 key. By pressing the +/
key, the 
polarity of the value can be alternated.  Press 
the 
ENTER
 key to store the setting value into 
memory. After this the display returns to selection 
of the Function number. 
Analog output terminal
 
DC 7V - 10V 
AC adapter Jack 
+
-
 
 
19. 4-20mA Analog Output (OP-07)
Output current 
Weight
20mA
4mA
f32 a 
f33 a
f32 b 
f33 b