AD-4405 Weighing Indicator
 
 
20.
20.
20.
20.     Current Loop Output (OP
Current Loop Output (OP
Current Loop Output (OP
Current Loop Output (OP-
-
-
-08)
08)
08)
08)    
  Replacing the RS-232C interface with this option, current loop output, 3-relay outputs 
and 1-control input can be used with the RS-232C interface of this option. 
  RS-232C functions and pin connections are the same as the RS-232C interface 
described in “16. RS-232C Interface”. 
  The control input can control the indicator from an external terminal just like the front 
panel key operations. 
  Set the external control function at f15 of the F-Functions. 
  When connecting the function pin and the common pin, the indicator makes the action. 
  Keep a signal width of more than 100 ms for the On-time and Off-time. 
 
   
ON(Make) 
More than 100 ms  More than 100 ms 
 
  OFF(Break) 
  The relays output the result of comparison. 
Solid-state-relay 
Maximum voltage 
DC50V 
Maximum current 
DC100mA 
Maximum resistance  8
9
 
  Pin connections (DIN 8pin connector) 
Function 
Pin No.  Signal name
Direction
Description 
HI 
Output 
Relay output HI 
OK 
Output 
Relay output OK 
LO 
Output 
Relay output LO 
Relay 
output 
8 COM(out) - 
Relay 
common 
terminal 
CL1 
Output 
Current loop output 1 
Current 
loop 
CL2 
Output 
Current loop output 2 
EXT3 
Input 
Control input 3 (f15) 
Control 
input 
COM(in) 
Control input common 
  Adaptable connector 
JA-TCP0586 (of accessory) 
Circuits
Circuits
Circuits
Circuits    
 
 
Current loop output 
HI
OK
LO
COM(out)
EXT3
+7 to 9V
680
9
COM
(in) 
DIN connector 
of OP-08 
Control input Relay output
20. Current Loop Output (OP-08)