Omega 3200 User Manual

Page of 26
M-4271/0707, pg. 15 of 26 
 
Q
1
 / Q
2
  = K
1
 / K
2
 
 
Q
1
 is the flow rate of the new gas 
Q
2
 is the flow rate of the original calibration gas 
K
1
 is the K factor of the new gas 
K
2
 is the K factor of the original calibration gas 
 
Q
1
 = (K
1
 / K
2
) Q
2
   
 
 
 
 
If K
2
 is larger than K
1
 then linear results will only be achieved if the unit 
does not exceed 5(K
1
/ K
2
)VDC for the full scale output. 
 
Example 1 
For a 0-200sccm unit calibrated for air the flow at 5.0VDC would be 
200sccm. The K factor for air is 1. If the unit is used with Helium (K factor 
1.454 relative to air) then the flow at 5VDC (i.e. the maximum flow) would 
be (1.454/1)200 = 290.8 sccm  
 
Example 2 
For a 0-10.0 l/min unit calibrated for Argon the flow at 5.0VDC would be 
10.0l/min. The K factor for Argon is 1.45. If the unit is used with Carbon 
Dioxide (K factor 0.74) then the flow rate 5.0VDC would be 
(0.74/1.45)10.0 = 5.10l/min 
 
The accuracy of readings using K factors is not as good as that achieved 
for the calibration gas. The accuracy obtained (typically ±3% for K factors 
similar to the calibration gas) depends on the gas being used and the flow 
rate. 
 
For a list of common K Factors see Section J. 
 
4.  Changing The Flow Rate Set-Point (Using An External Voltage 
Source) 
 
The required flow rate is selected by adjusting the set-point voltage. The 
normal control signal voltage is 0-5VDC with 0VDC corresponding to zero 
flow and 5VDC being equivalent to the maximum rated flow of the unit. 
This input is linear and scaleable allowing different flow rates within the 
range of the unit to be selected. For example: 
 
 
 
For a flow range of 0-500sccm: 
A 5 VDC Input Signal would correspond to a flow rate of 500sccm  
If a flow rate of 300sccm were required then the set-point would 
be: 
(300 ÷ 500) × 5 = 3.0VDC