7. Detailed Description
Note: Although this feature activates the general cut-off, it is
not a dedicated ”safety-function” of the type listed
above. When the error message shown above is dis-
played, the sensor must be exchanged by service per-
sonnel. 
Restart after power failure
MultiTemp III automatically restarts after a power failure.
Make sure that this automatic restart will not cause a dange-
rous situation to arise. 
The unit can be restarted manually instead by pressing 
(see Section 7.5).
MultiTemp III is factory set to normal mode. Normal mode
means that the unit automatically re-starts after power failu-
re. As most modern electrophoresis power supplies automati-
cally restart after short power failures, we strongly recom-
mend that you leave MultiTemp III set in its normal mode.
However, if automatic re-start is inconvenient, the unit can
be restarted manually by setting Operation mode 8 to ON,
follow the procedure below:
Setting mode 8 to ON
1. Press          simultaneously with the mains switch. Then
press 
again until ”mode 8” is selected.
2. 
Press 
to select ”ON” instead of ”OFF”.
(”OFF” is factory set) 
3. 
Confirm by pressing 
when the display is flas-
hing. 
4. 
Wait until the display shows the actual temperature.
5. 
Switch off the unit using the mains switch. 
The operation mode is now changed to ON. 
7.5 Manual 
restart in 
operation 
mode 8
20