Trendnet 18-1106-33 User Manual

Page of 27
8. Technical specifiations
8. Technical
specifications
Technical specifications (ambient temperature 20 °C)
Working temperature range
-10 to 90 °C
Temperature control constancy at 20 °C
< ± 0.1 °C
Range of temperature limit switch
20 to 95 °C
Heater capacity,
230V
800 W 
115V
800 W
Cooling capacity,
115 V
at 20 °C 
240 W
at 5 °C
200 W
Cooling capacity, 
230 V 
at 20 °C 
265 W
at 5 °C
215 W
Compressor coolant
R 134 a
Bath volume
3 litres
Bath opening/depth
130x100/100 mm
Pump capacity, pressure, max.
0.32 bar
Pump capacity, flow rate, max.
12 l/min
Noise level at 1 m in front of the unit
<63 dB(A)
Mains requirements
95 to 132 V/ 60 Hz 
95 to 125 V/ 50 Hz 
198 to 253 V/ 50 Hz
Power consumption
1,200 W
Ambient operating humidity
Relative humidity maximum 
80% at 31 °C; maximum 50% 
at 40 °C
Ambient operating temperature 
4 to 40 °C
Ambient operating pressure
Maximum altitude of 2,000 m
Dimensions (H x W x D)
480 x 220 x 420 mm
Weight
28 kg