Agilent Technologies 81480A User Manual

Page of 194
Cleaning Information
What do I need for proper cleaning?
174
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, 
Sixth Edition
Furthermore, the power density may burn dust into the fiber and 
cause additional damage (for example, 0 dBm optical power in a single 
mode fiber causes a power density of approximately 16 million W/m
2
). 
If this happens, measurements become inaccurate and non-repeatable.
Cleaning is, therefore, an essential yet difficult task. Unfortunately, 
when comparing most published cleaning recommendations, you will 
discover that they contain several inconsistencies. In this section, we 
want to suggest ways to help you clean your various optical devices, 
and thus significantly improve the accuracy and repeatability of your 
lightwave measurements. 
What do I need for proper 
cleaning?
Some Standard Cleaning Equipment is necessary for cleaning your 
instrument. For certain cleaning procedures, you may also require 
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following 
standard equipment:
Dust and shutter caps
All of Agilent Technologies’ lightwave instruments are delivered with 
either laser shutter caps or dust caps on the lightwave adapter. Any 
cables come with covers to protect the cable ends from damage or 
contamination.