Agilent Technologies 81480A User Manual

Page of 194
Cleaning Information
What do I need for proper cleaning?
176
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, 
Sixth Edition
Soft tissues
These are available from most stores and distributors of medical and 
hygiene products such as supermarkets or chemists' shops. 
We recommend that you do not use normal cotton tissues, but multi-
layered soft tissues made from non-recycled cellulose. Cellulose 
tissues are very absorbent and softer. Consequently, they will not 
scratch the surface of your device over time.
Use care when cleaning, and avoid pressing on your optical device 
with the tissue. Pressing too hard may lead to scratches on the surface 
or misalignment of your device. Just rub gently over the surface using 
a small circular movement.
Use only clean, fresh soft tissues and never apply them twice. Any dust 
and dirt from the air which collects on your tissue, or which has 
gathered after initial cleaning, may scratch and pollute your optical 
device.
Pipe cleaner
Pipe cleaners can be purchased from tobacconists, and come in 
various shapes and sizes.The most suitable one to select for cleaning 
purposes has soft bristles, which will not produces scratches.
There are many different kinds of pipe cleaner available from tobacco 
nists.
The best way to use a pipe cleaner is to push it in and out of the device 
opening (for example, when cleaning an interface). While you are 
cleaning, you should slowly rotate the pipe cleaner.
Only use pipe cleaners on connector interfaces or on feed through 
adapters. Do not use them on optical head adapters, as the center of a 
pipe cleaner is hard metal and can damage the bottom of the adapter.
Your pipe cleaner should be new when you use it. If it has collected 
any dust or dirt, this can scratch or contaminate your device.
The tip and center of the pipe cleaner are made of metal. Avoid 
accidentally pressing these metal parts against the inside of the device, 
as this can cause scratches.
Compressed air
Compressed air can be purchased from any laboratory supplier.
It is essential that your compressed air is free of dust, water and oil. 
Only use clean, dry air. If not, this can lead to filmy deposits or 
scratches on the surface of your connector. This will reduce the 
performance of your transmission system.