Agilent Technologies 81480A User Manual

Page of 194
What do I need for proper cleaning?
Cleaning Information
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
175
We suggest these protective coverings should be kept on the 
equipment at all times, except when your optical device is in use. Be 
careful when replacing dust caps after use. Do not press the bottom of 
the cap onto the fiber too hard, as any dust in the cap can scratch or 
pollute your fiber surface.
If you need further dust caps, please contact your nearest Agilent 
Technologies sales office. 
Isopropyl alcohol
This solvent is usually available from any local pharmaceutical 
supplier or chemist's shop.
If you use isopropyl alcohol to clean your optical device, do not 
immediately dry the surface with compressed air (except when you 
are cleaning very sensitive optical devices). This is because the dust 
and the dirt is solved and will leave behind filmy deposits after the 
alcohol is evaporated. You should therefore first remove the alcohol 
and the dust with a soft tissue, and then use compressed air to blow 
away any remaining filaments.
If possible avoid using denatured alcohol containing additives. 
Instead, apply alcohol used for medical purposes. 
Never drink this alcohol, as it may seriously damage to your health.
Do not use any other solvents, as some may damage plastic materials 
and claddings. Acetone, for example, will dissolve the epoxy used with 
fiber optic connectors. To avoid damage, only use isopropyl alcohol.
Cotton swabs
We recommend that you use swabs such as Q-tips or other cotton 
swabs normally available from local distributors of medical and 
hygiene products (for example, a supermarket or a chemist's shop). 
You may be able to obtain various sizes of swab. If this is the case, 
select the smallest size for your smallest devices.
Ensure that you use natural cotton swabs. Foam swabs will often leave 
behind filmy deposits after cleaning.
Use care when cleaning, and avoid pressing too hard onto your optical 
device with the swab. Too much pressure may scratch the surface, and 
could cause your device to become misaligned. It is advisable to rub 
gently over the surface using only a small circular movement.
Swabs should be used straight out of the packet, and never used twice. 
This is because dust and dirt in the atmosphere, or from a first 
cleaning, may collect on your swab and scratch the surface of your 
optical device.