Agilent Technologies E4445A User Manual

Page of 664
442
Chapter 5
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
most useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth
medthod is preferred for other signals.
Key Path:
Meas Setup
Key Notes:
When Method is set to RBW, neither Noise Correction nor RRC Filter are
available.
State Saved:
Saved in instrument state.
Factory Preset: IBW
Range:
IBW|RBW
History:
Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:MCPower:METHod IBW|RBW
[:SENSe]:MCPower:METHod?
Example:
MCP:METH RBW
MCP:METH?
5.19.8
Power Ref
Enables you to set the multi-carrier power reference to automatic or manual. When set to
automatic, the carrier power result reflects the measured power value in the selected
reference carrier (
Meas Setup
,
Carrier Setup
,
Ref Carrier
). When set to manual, the result is
referenced to the last measured value, or you may specify the reference for the
multi-carrier power measurement. Relative values are displayed, referenced to the “Power
Reference” value.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: Off/Measured power in the reference carrier.
Terminators:
dBm
Default Terminator: dBm
History:
Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:MCPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:CARRier:AUTO[:STATe]?
[:SENSe]:MCPower:CARRier[:POWer]<dBm>
[:SENSe]:MCPower:CARRier[:POWer]?