Agilent Technologies E4445A User Manual

Page of 664
444
Chapter 5
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Remote Command:
[:SENSe]:MCPower:LIMit[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:LIMit[:STATe]?
Example:
MCP:LIM ON
MCP:LIM?
5.19.10
RRC Filter
Pressing
RRC Filter
turns the Root Raised Cosine filter on or off for the carriers and all
adjacent channels. The RRC filter is the type specified in the 3GPP W-CDMA standards,
with roll-off (
α) =
parameter (defined below).T is set to 1/Carrier Integ BW (for
carriers) or 1/Offset Integ BW (for Offsets).
Key Path:
Meas Setup
,
More
Key Notes:
This key is not available when
Meas Setup
,
Method
(RBW) is selected.
State Saved:
Saved in instrument state.
Factory Preset: Off
History:
Added with firmware revision A.02.00
Remote Command:
[:SENSe]:MCPower:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:MCPower:FILTer[:RRC][:STATe]?
Example:
MCP:FILT 1
MCP:FILT?
5.19.11
Filter Alpha
Press
Filter
to input the alpha value for the RRC Filter. This parameter is only available
when
3GPP W-CDMA
or
None
has been selected as the radio standard from the Mode Setup
menu.
Key Path:
Meas Setup, More
State Saved:
Saved in instrument state.
Factory Preset: 0.22 when W-CDMA is selected, otherwise Off.
Range:
Off/On
History:
Added with firmware revision A.03.00.