A&D GP-100K User Manual

Page of 98
 
 25
7. CALIBRATION 
7-1 Calibration Group 
The GP series balance has the following modes as a calibration group.   
Calibration 
  Automatic self calibration (calibration due to changes in temperature) 
    
  Calibration using the internal mass (one-touch calibration) 
    
  Calibration using an external weight 
Calibration test 
 Calibration test using an external weight (Calibration test does not 
perform calibration.) 
Correction of the internal mass value 
 
Terms 
The following terms are defined as follows: 
Internal mass   
=  Built-in calibration weight 
External weight    =  A weight that you have. Referred to as a calibration weight when used for calibration. 
Calibration weight   =  A weight used for calibration 
Target weight   
=   An external weight used for calibration test 
 
Caution 
  Calibration adjusts the balance for accurate weighing.   
  Besides periodic calibration and before each use, perform calibration when: 
  • the balance is installed for the first time. 
  • the balance has been moved. 
  • the ambient environment has changed. 
  Do not allow vibration or drafts to affect the balance during calibration. 
  To output the data for GLP using the RS-232C interface, set “GLP output (info)” of “Data 
output (dout)”. For details, refer to “9. FUNCTION TABLE”. Time and date are added to the 
GLP report. If the time or date is not correct, adjust them. For details, refer to “9-9 Clock and 
Calendar Function”.   
  Calibration test is available only when “GLP output (info)” of “Data output (dout)” is set to “1” or “2”, 
  The calibration and calibration test data can be stored in memory. To store them, set “Data 
memory (data)” to “3”. For details, refer to “11. DATA MEMORY”. 
 
Caution on using an external weight 
 The accuracy of an external weight will influence the accuracy of weighing. Select an 
appropriate weight as listed below:
                                                                                   
 
Model 
Usable calibration weight 
Adjustable range 
GP-12K 5 
kg, 
10 kg 
GP-20K 10 
kg, 
20 kg 
GP-30K/GP-30KS 
20 kg, 30 kg 
GP-32K/ GP-32KS 
20 kg, 30 kg 
GP-40K 
20 kg, 30 kg, 40 kg 
 
 
-1.5 g  to  +1.5 g 
 
   
GP-60K/ GP-60KS 
30 kg, 40 kg, 50 kg, 60 kg 
-15 g  to  +15 g 
GP-61K/ GP-61KS 
30 kg, 40 kg, 50 kg, 60 kg 
-1.5 g  to  +1.5 g 
GP-100K/GP-100KS 
60 kg, 80 kg, 100 kg 
GP-102K 
60 kg, 80 kg, 100 kg 
-15 g  to  +15 g 
The calibration weight in bold type: factory setting 
The calibration weight value can be adjusted within the range above.