10.0  RELAY
 
INTERFACE 
 
 
The AE 402 indicator is supplied with drivers to control external relays.  The drivers 
could be used to control a number of different relays depending upon the users 
need.    The relay drivers are isolated outputs requiring the use of an external power 
supply for the relays.    Also see section 3.4.3. 
 
Contact Adam Equipment or your supplier for a relay interface that is compatible 
with the relay drivers, however other interfaces can be used as long as the following 
conditions apply. 
 
Connections to the drivers: 
 
 
 
Connections are made to the PCB, Connector P1.   The connector is a spring 
activated type, simply press on the top of the connector and insert the wire. 
 
Do not exceed the safe limits of voltage or current of 24VDC or 15ma per output. 
 
Depending upon the application it may be necessary to use a small relay to drive 
larger relays, or to provide additional protection to prevent electromagnetic 
interference (diodes as shown above) to this or other machinery.