11.0  CALIBRATION 
 
 
The scale can be calibrated using the following procedure.     To enter this procedure 
it is necessary to use Func 4 accessible using the [Func/C] key as described in section 
7.4   or using the password access as described in section 12. 
 
The AE 402 scales calibrate using either metric or pound weights, depending on the 
weighing unit in use before calibration.   The display will show either "kg" or "lb" to 
identify the weights expected. 
 
 
PROCEDURE 
 
•  Enter the calibration section using Func 4, C8 CAL or using the password 
as described in section 12.    
 
•  The display will show "unLoAd". 
 
•  Remove any weight from platform, when the stable symbol is on press 
[Z/T]. 
 
•  The  display  will  show  "Ld"  the  “0000XX”  to  show  the  last  calibration 
weight used.    Place this calibration weight on the scale. Press the [Z/T] 
key.      If  the  weight  you  put  on  the  scale  does  not  match  the  value 
displayed,  press  the  [Func/C]  key  to  clear  the  vale  then  use  the 
[Unit/
]key  and  [Pcs/
]  key  to  set  the  correct    value.    When  it  is 
correct press [Z/T]. 
 
•  If the calibration is acceptable the scale will run a self-test during which 
the calibration weight should be removed.  If an error message “FAiL L” 
is  shown  try  calibration  again  as  a  disturbance  may  have  prevented  a 
successful calibration.   
 
After  calibration  the  scale  should  be  checked  to  verify  the  calibration  is  correct.  If 
necessary repeat calibration, make sure that the scale is stable before accepting any 
weight.