3.5  WEIGHING 
 
  Place the sample on the scale platform. The weight will be 
displayed in the selected Unit.  
 
  Wait until the stable indicator is displayed for accurate 
weighing. 
 
 
3.6  CHECK WEIGHING 
 
Check-weighing  is  a  procedure  to  sound  an  alarm  when  the  weight  of 
the  sample  on  the  scale  meets  or  exceeds  the  pre-set  Low  and  High 
values stored in the memory. 
 
To set the High and Low values, follow the procedure given below- 
 
 
Press  the  [High/Low] key.  The  display  will  show  “
000.00  HI
” 
with  the  first  digit  flashing  for  setting  the  High  value.  To  set 
your limit, use the [Unit] key to advance the flashing digit, the 
[Tare] key to decrease the value of the flashing digit and the 
[Zero] key to increase the value of the flashing digit .
 
 
  The  display  will  now  show  “
000.00  LOW
”  with  the  first  digit 
flashing  for  setting  the  Low  value.  To  set  your  limit,  use  the 
[Unit]  key  to  advance  the  flashing  digit,  the  [Tare]  key  to 
decrease the value of the flashing digit and the [Zero] key to 
increase the value of the flashing digit . Press [Mode] again 
to  confirm  the  value  and  the  display  will  return  to  normal 
weighing mode. 
 
  Once  the  check  weighing  Limits  are  set  and  the  alarm 
function is active, a beep will sound slowly when the sample 
weight is below the Low Limit. The beep will sound quickly if 
the weight is above the High Limit. If the sample weight is in 
between the Limits, the beep does not sound.