K-jump Health Co. Ltd. TEMPONITOR User Manual

Page of 6
8
9
Error Message
Specifications
Power Source
6V ; 1.5Vdc AAA size battery X4
DC
Measuring range
90.0 F ~ 109.9 F,(32.0 C ~ 43.9 C)
Resolution
0.1 F (
0.1 C)
Accuracy
0.2°F (±0.1°C): 93.2~107.6°F (32.0 ~
42.0°C).
0.4°F (±0.2°C): other range.
Operation Frequency
13.56MHz
Operation Range
within 4cm diameter circle with distance
within 3cm. 
Memory
automatic display last measurement
Low battery
Replace the batteries if the low battery 
indicator appears.
Power OFF
Automatically after 1 minutes of non use.
Operating environment
+60.8 F~+104.0 F(
16.0 C ~ 
40.0 C),at
95%RH
Storage environment
-68.0 F~+122.0 F(
20.0 C ~ 
50.0 C),at
95%RH
       or
No display
Error Marks
Correction
The temperature 
measured was higher
than the specified
operating range.
Please measure
again.
The temperature
measured was lower
than the specified
operating range.
Please measure 
again.
Low Battery Indication
Replace four 1.5V
size AAA batteries
System Failure
The device is either
out of function or 
unable to store
memory data. Check
with the dealer.
Transmission Failure
1. Re-position the
reader to closer
distance.
2. Replace a new patch 
due to the one is 
malfunction.
3. Replace the batteries.