10.4.2 Start up
Problem
Cause
Action
MobiMed 300c does 
not start (green diode 
is not lit).
300c has not been 
turned on properly.
Hold On/Off button on 300c pressed for at least 2 
seconds.
Battery low in 300c 
Connect external power to charge the batteries in 
300c. Make sure that the two indicators for external 
power supply connected and battery charging are lit. 
See “Units in the system” on page 5. Try pressing start 
button after a few minutes of charging.
MobiMed 300m does 
not start.
300m has not been 
started properly.
Press the Monitor on button on the 300m.
Battery low in 300m. Connect external power to charge the batteries in the 
300m. Make sure that the 300m receives external 
power. Try pressing start button after a few minutes of 
charging.
MobiMed 300m can 
not establish contact 
with 300c.
The 300c is not 
turned on.
Check that the 300c is turned on (green diode is lit).
Bluetooth
 connection 
is not working.
Check that the 300c is within radio range of 300m 
(approx. 5 metres).
Press ID button on 300c to display a pop-up menu 
with the 
Bluetooth
 address. If it does, it means that the 
300m has contact with the 300c.
Check that right 
Bluetooth
 address is stated in Unit 
connection settings, see “Unit connection settings” on 
page 63. Compare with 
Bluetooth
 address on product 
label. Otherwise connect to appropriate 300c.