Ortivus AB 410000409 User Manual
71
Appendix 10
10.4.2 Start up
Problem
Cause
Action
MobiMed 300c does
not start (green diode
is not lit).
300c has not been
turned on properly.
Hold On/Off button on 300c pressed for at least 2
seconds.
Battery low in 300c
Connect external power to charge the batteries in
300c. Make sure that the two indicators for external
power supply connected and battery charging are lit.
See “Units in the system” on page 5. Try pressing start
button after a few minutes of charging.
MobiMed 300m does
not start.
300m has not been
started properly.
Press the Monitor on button on the 300m.
Battery low in 300m. Connect external power to charge the batteries in the
300m. Make sure that the 300m receives external
power. Try pressing start button after a few minutes of
charging.
MobiMed 300m can
not establish contact
with 300c.
The 300c is not
turned on.
Check that the 300c is turned on (green diode is lit).
Bluetooth
connection
is not working.
Check that the 300c is within radio range of 300m
(approx. 5 metres).
Press ID button on 300c to display a pop-up menu
with the
Bluetooth
address. If it does, it means that the
300m has contact with the 300c.
Check that right
Bluetooth
address is stated in Unit
connection settings, see “Unit connection settings” on
page 63. Compare with
Bluetooth
address on product
label. Otherwise connect to appropriate 300c.